Use Instructions

3.3 Product installation and setup
ML6710i is not a permanently installed device i.e. it doesn’t require permanent (detachable by the use of
a tool only) electrical connection to supply mains. The device does not have or require a separate supply
mains switch and is isolated from the supply mains simply by pulling the plug from the mains socket.
Setup of the ML6710i is to be done at the treatment site by personnel authorized by the manufacturer
(Figure 4).
Figure 4. Example setup of ML6710i on an ophthalmic treatment station.
Left: Illustration of a physician (person on right) directing the laser treatment beam into the eye of the patient
(person on left) via a slit lamp and a contact lens on an ophthalmic table.
Right: A typical treatment station setup with ML6710i. 1: ML6710i laser device, 2: iPad (GUI) with wireless
connection to ML6710i, 3: Modulight cloud service accessed via GUI, 4: Foot switch for activating laser
emission, 5: ML-SLA accessory for shaping the treatment beam, 6: Slit lamp/ophthalmic microscope,
7: Ophthalmic contact lens to deliver treatment beam into the retina of the patient, 8: Ophthalmic table.
In case a calibrated treatment set-up (including laser, SLA and slit lamp) is disassembled or transported
to another location the treatment station should be re-calibrated by service personnel authorized by
the manufacturer. Please contact the manufacturer for further instructions or agreeing a service visit.
MANDATORY
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
WARNING
ML6710i Instructions For Use
20 Product delivery package and installation