Use Instructions
Table 3. Electrical characteristics for ML6710i.
Parameter Range Unit Comment
Input voltage 15 V ± 5%
Input power ≤ 60 W
Current consumption ≤ 4 A
Input voltage of power supply 100–240 VAC
Input frequency of power supply 50–60 Hz
Max current consumption
1.5
0.7
A
115 VAC
230 VAC
Bluetooth Low Energy frequency of reception & transmission 2400–2483.5 MHz Bandwidth 2 MHz
Effective radiated power 6 dBm Nominal
Parameter Unit Min Max Comment
Operating temperature °C 15 30
Operating humidity %RH 0 70
Non-condensing operating environment
required
Operating atmospheric
pressure
kPa 70 106
Storage and transportation
temperature
°C -20 +50
Storage and transportation
humidity
%RH 0 90 Non-condensing environment required
Storage and transportation
atmospheric pressure
kPa 70 106
Table 4. Ambient conditions for ML6710i.
2.4.2 Ambient conditions
The device is designed to be operated in typical hospital or clinic environments, with the ambient environment
characteristics specified in Table 4. Exclusions regarding electromagnetic compatibility is listed in chapter 5.
The device should be allowed to return to normal operating conditions after transportation or storage prior to use.
Revision 9
15The ML6710i