Use Instructions

2.4 Specification
2.4.1 Electro-optical and electrical performance
The ML6710i has a specific usable output power range, which limits the available spot size
depending on the selected contact lens magnification and irradiance. Spot diameter is
limited by the SLA properties. Settings outside these ranges are indicated on the GUI.
Parameter Symbol Unit Min Typ Max Comment
Target wavelength
1
λ
nm 689 90% power within 689 ± 3 nm
SLA output power
2
P
SLA
mW 400
ML-SLA output power being within 20% of set
value
Stability
2
% ±10
At maximum power over one treatment cycle,
from 5 s after start
Aiming beam power P
AB
mW 0.02 2.5 Class 2 Laser @ 635 ± 10 nm wavelength
Spot diameter
3
mm 1.0 5.0
Without contact lens. All values full width at half
maximum (FWHM) with ±10% accuracy
Contact lens
magnification
0.5 2.5
Multiplied with Spot diameter to achieve actual
spot at retina
Irradiance E mW/cm
2
50 600 800
Depends on indication. Limited by selected
contact lens.
Dose H
e,o
J/cm
2
5 50 95
Please consult the prescribing information of the
drug for recommended light dose
1
The wavelength is fixed by the manufacturer and is automatically controlled by the laser device
2
The output power is automatically adjusted and stabilized by the device based on the set treatment parameters
3
In focus after SLA
Table 2. Electro-optical characteristics at 59–86 °F (15–30 °C).
ML6710i Instructions For Use
14 The ML6710i