User manual
APPENDIX B – MAINTENANCE
21
ENGLISH
Osmotic membrane
See Figure Q
NOTE
AFTER REPLACING THE OSMOTIC MEMBRANE AND THE PURGING OF THE
HYDRAULIC CIRCUIT
, REPLACE ALSO THE ACTIVE CARBON FILTER.
1. Turn off MILLDROP and remove the mains cable.
2. Proceed as with the resin filters and remove the whole filter assembly (A) from its
bearings to access the osmotic membrane assembly (B).
3. Disconnect the pipes Ø8 (1) and (2) from the fittings and replace the osmotic
membrane.
4. Reconnect the pipes respecting the original position.
5. Replace the active carbon filter as described on the next page.
NOTE
T
O DISCONNECT THE PIPES FROM THE FITTINGS, PULL OUTWARDS WHILE KEEPING
PRESSED THE FITTING LOCK RING
.
Figure Q
6. Modify the wire connection on the terminals board as described in the procedure for
the purging of the hydraulic circuit (see Fig. F), disconnect the pipe from the input
resin filter fitting (see Fig. G) and perform the purging for 10 minutes.
7. Check for no leak on the equipment.
8. Restore the hydraulic and the electric connections, and replace the cover.
9. Check the quality of output demineralized water (see table).
A
B
1
2
1