User manual
FIRST STARTUP
15
ENGLISH
FIRST STARTUP
MILLDROP
CONNECTED TO
THE STERILIZER
SERIES
MILLENNIUM
Once the MILLDROP has been properly installed, proceed with the following additional
startup operations and checks according with the type of use.
Turn on the demineralizer and then the sterilizer.
Ended the initial test of the sterilizer, enter the SETUP menu, select “Filling Options” and
the mode “AUTOMATIC” (see the Operating Manual of the sterilizer).
Check that demineralizer starts to work as the sterilizer’s automatic feeding pump starts,
and stops at the same time as the pump.
NOTE
THE EQUIPMENT WILL STOP AUTOMATICALLY IF THE DELIVERY OF WATER LASTS
UNINTERRUPTEDLY FOR MORE THAN
20 MINUTES. TO RESTART THE MILLDROP
OPERATION TURN OFF AND ON THE UNIT AGAIN
.
MILLDROP
CONNECTED TO
THE FEEDING
TANK
The demineralizer starts immediately as turned on.
Reached the maximum level in the tank (signaled through the level sensor), the
demineralizer stops.
Turn on the sterilizer.
Ended the sterilizer initial test, enter the SETUP menu, select “Filling Options” and the
mode “AUTOMATIC” (see the Operating Manual of the sterilizer).
If the water level into the sterilizer’s internal tank is lower the minimum, the automatic
feeding pump starts.
In this case, the demineralizer will start a few seconds after the start of the sterilizer’s pump
(i.e. as the tank’s float, by lowering, supplies the start/stop command to the demineralizer),
and will stop as the tank’s float detects again the maximum threshold.
NOTE
THE EQUIPMENT WILL STOP AUTOMATICALLY IF THE DELIVERY OF WATER LASTS
UNINTERRUPTEDLY FOR MORE THAN
20 MINUTES. TO RESTART THE MILLDROP
OPERATION TURN OFF AND ON THE UNIT AGAIN
.
In case of manual water drawing also from the tank’s tap, ensure that the water remaining
in the tank is sufficient to warrant the need of sterilizer(s) connected to the tank,
considering also the time necessary to restore a sufficient level in the tank. In substance:
- one sterilizer connected, the remaining water must be at least 2 liters .
- two sterilizers connected, the remaining water must be at least 4.5 liters.
- three sterilizers connected, the remaining water must be at least 7 liters.
MILLDROP
CONNECTED TO
THE MANUAL
DRAWING TAP
Press the tap’s button to start the water delivery.
Press again the tap’s button to stop the water delivery.
NOTE
THE EQUIPMENT WILL STOP AUTOMATICALLY IF THE DELIVERY OF WATER LASTS
UNINTERRUPTEDLY FOR MORE THAN
20 MINUTES. TO RESTART THE MILLDROP
OPERATION TURN OFF AND ON THE UNIT AGAIN.