User manual

INSTALLATION
10
Purging the internal
hydraulic circuit
(osmotic membrane)
See Figure F and G:
1. Remove the screws and the lid (see Figure A) to access the electronic board (see
Figure F)
Electronic board type 1
:
- Remove the protection panel of the electronic board, unplug the signal terminal
board.
- Remove the blue wires (coming from the relay board for the sterilizer automatic
filling interface) from pins 5-6 of the terminal board.
- Make a “jumper” between pins 5-6 and fit the terminal board on the electronic
board.
Electronic board type 2
:
- Remove (if present) the wires from pins 3-4 of the terminal board, and make a
“jumper”.
Figure F
2. Disconnect the pipe from the input resin filter fitting, extend it by using the silicone
pipe supplied and take its end in a washbasin (figure G).
Figure G
Red wires
Relay board
Terminal board
Blue wires
Jumper
Jumper
Electronic board type 1 Electronic board type 2
toward a washbasin
Disconnect and
extend with a
silicone pipe