User manual

INSTALLATION
9
ENGLISH
HYDRAULIC
CONNECTIONS
Initial installation
See Figures D, E:
1. Use the kit supplied and install the draining bracket onto the washbasin’s siphon
where you wish to install the MILLDROP draining pipe; carry out an hole Ø 6 mm in
the siphon across the bracket’s fitting.
2. Insert the red pipe in the ring nut , plug it in the bracket’s fitting, and fasten the ring
nut.
drill Ø6
red pipe
Siphon
Bracket
Figure D
3. Cut the red pipe long enough to reach the draining connection B on the rear panel of
the demineralizer, and plug-in the pipe.
NOTE
THE PIPING MUST NOT BE BENT, SQUASHED OR HAVE A PARTICULARLY TORTUOUS
ROUTE
, OTHERWISE PERFORMANCE WILL BE QUALITATIVELY VERY LOW OR, WHAT
IS WORSE
, THE DEMINERALIZER ITSELF SHOULD MALFUNCTION.
4. Install a tap (not supplied) provided with male connector ¾” on the water outlet and
near the demineralizer. Check the pressure value of the water works (see Appendix
A).
5. Connect the feeding pipe (blue) between the tap outlet and the water inlet A on the
rear panel of the demineralizer (check the presence of gaskets), fasten both the
pipe’s ring nuts.
6. Open the tap and check for no leaks.
7. Remove the plug from the outlet C on the rear panel of the demineralizer and plug-in
the pipe fitting.
8. Mount the reinforced silicone pipe on the pipe fitting, tight by clamps.
9. Place the opposite end of the silicone pipe inside a wash-basin.
Figure E
Gasket
Blue pipe
Pipe-union
Red pipe
Draining siphon
Wash-basin or
external tank
Reinforced
silicon
p
i
p
e
TAP
Clamp