User manual

PRODUCT
5
ENGLISH
FRONT PANEL
REAR PANEL
C
B
S
FUNCTIONING
The MILLDROP operation is very easy and automatic in any use.
1) Directly connected to a Millennium series sterilizer, it operates at the start of the
sterilizer’s water feeding pump, and until the pump stops.
2) Connected to the external tank, it starts working as the water level in the tank goes
down the maximum threshold, and stops as it is again reached.
If continually working for a time exceeding 20 minutes, MILLDROP shall stop
automatically (see Note).
3) Connected to the drawing tap (provided with electric device), it starts working as you
press the tap’s button and stops pressing it again.
If continually working for a time exceeding 20 minutes, MILLDROP shall stop
automatically (see Note).
NOTE
TO RESTART THE DEMINERALIZER FROM A STOP STATUS, TURN IT OFF AND ON
AGAIN
.
Green LED – good quality of the outlet water
Yellow LED– sufficient quality
Red LED– poor quality
Red LED– lack of inlet water or low pressure
Red LED– internal water leak
Connector of the water inlet solenoid valve
Solenoid valve filter
Outlet of the permeation product
Start/Stop interface connector
220Vac input – Fuses – Mains switch
Draining connector