APPENDIX E – ALARMS
87
CODE POSSIBLE CAUSE PROPOSED SOLUTION
A 111
Incorrect
connection of the
temperature sensor (sterilization
chamber) to the connector.
Contact the Technical Support Department
(see Appendix Z
).
Temperature sensor short circuit
(sterilization chamber).
A 112
Incorrect
connection of the
temperature sensor (steam
generator) to the connector.
Temperature sensor short circuit
(steam generator).
A 113
Incorrect
connection of the
temperature sensor (heating
element) to the connector.
Temperature sensor short circuit
(heating element).
A 114
Incorrect
connection of the
temperature sensor (chamber wall)
to the connector.
Temperature sensor short circuit
(chamber wall).
A 200
Intervention of the steam generator
safety thermostat.
Manually rearm
the thermostat(s) located on the back of the device (see
the Chapter,
Product Introduction”).
Unscrew the black plastic protection cap, press the red button
until you
hear a click and replace the cap.
Turn-off (RESET) and then turn-on the device.
If the problem persists contact the Technical Support Department
(see Appendix Z
).
Intervention of the heating element
safety thermostat.
Heating or steam generator heating
element malfunction.
A 250
Presence of water or condensate in
the sterilization chamber.
Perform RESET following the instructions.
Carefully dry the inside of the sterilization chamber and start the program
again.
Do not
put material impregnated with water, or liquids in general, in the
chamber.
Drain filter of the sterilization
chamber obstructed.
Clean or replace the drain filter
(See Appendix C
Maintenance”).
Air leaking through the gasket.
Perform RESET following the instructions.
Carefully clean the gasket with a clean cotton cloth dampened with
water.
Start the program again.
Vacuum pump broken.
Contact the Technical Support Department (see Appendix Z
).
Problem in the plumbing circuit.
A 251
Water injection pump malfunction.
Contact the Technical Support Department (see Appendix Z).
Problem in the plumbing circuit.
Intervention of the steam generator
safety thermostat.
See A200
If the problem persists contact the Technical Support Department
(see Appendix Z
).
Heating element safety thermostat
intervened.
Heating or steam generator heating
element malfunction.