 APPENDIX E – ALARMS 
  83 
ALARM CODES 
The list
 of alarm codes and, consequently, the messages displayed on the LCD and relative RESET mode, is as follows: 
CODE  ALARM DESCRIPTION  LCD INDICATION  RESET MODE 
ERRORS (category E) 
E 000 
Blackout BLACK-OUT 
Press key 
(> 3 seconds) 
E 010 
Door open  DOOR OPEN 
E 020 
Exceeded timeout for activating door lock 
system (closing) 
DOOR UNLOCKED 
E 021 
Exceeded timeout for activating door lock 
system (opening) 
DOOR LOCKED 
E 030 
Water in the fill tank at minimum (MIN) level  WATER MIN 
E 031 
Water in the drain tank at maximum (MAX) level  EXHAUST MAX 
E 041 
Filling the tank too frequently 
(automatic filling) 
FILLING PROBLEM 
E 900 
Vacuum Test failed 
(during the LEAKAGE PHASE) 
TEST FAILED 
E 901 
Vacuum Test failed 
(during the WAITING PHASE) 
TEST FAILED 
E 902 
Vacuum Test failed 
(vacuum pulse timeout exceeded) 
TEST FAILED 
E 999 
Manual cycle interruption  MANUAL STOP 
ALARMS (category A) 
A 022 
System door lock microswitches failed (OFF-
OFF) 
LOCKING PROBLEM 
Turning-off 
device 
A 023 
System door lock microswitches failed (ON-ON) LOCKING PROBLEM 
A 024 
System door lock microswitches failed (ON-OFF) LOCKING PROBLEM 
A 032 
Sensor-level problem  LEVEL PROBLEM 
A 040 
Failure to fill the tank 
(automatic filling) 
FILLING PROBLEM 
A 101 
PT1 broken 
(sterilization chamber) 
PTC BROKEN 
A 102 
PT2 broken 
(steam generator) 
PTC BROKEN 
A 103 
PT3 broken 
(heating element) 
PTC BROKEN 
A 104 
PT4 broken 
(sterilization chamber wall) 
PTC BROKEN 
A 111 
PT1 short-circuited 
(sterilization chamber) 
PTC SHORTCIRCUIT 
A 112 
PT2 short-circuited 
(steam generator) 
PTC SHORTCIRCUIT 
A 113 
PT3 short-circuited 
(heating element) 
PTC SHORTCIRCUIT 
A 114 
PT4 short-circuited 
(sterilization chamber wall) 
PTC SHORTCIRCUIT 
A200  Pre-heating not performed within the timeout 
(heating resistor problem). 
HEATING PROBLEM 