Specifications
APPENDIX B – PROGRAMS
61
PROGRAM SUMMARY TABLE - MILLENNIUM B
2
PROGRAM
DESCRIPTION
NOMINAL VALUES
BASIC PROGRAM
PARAMETERS
STERILIZABLE MATERIAL
NOTES
Temperature
(°C)
Pressure
(bar)
Holding time
(min)
Cycle type
(EN 13060: 2004)
Pre-vacuum
(F=fractionated; S=single)
Standard drying
(L=long; S=short)
Total cycle time
(average load ÷ max load)
Average consumption H
2
O
(ml/cycle)
Average energy consumption
(kWh/cycle)
TYPE
MAX TOTAL MASS
(kg)
MAX MASS
PER TRAY (kg)
MAX MASS
PER ARTICLE (kg)
134 °C B CYCLE 134 2,10 4 B F L 39÷42 675 0,8
Porous, unpackaged
material
1,25 0,40 0,30
For material and
instruments in
(single and
double)
packaging, we
recommend
using the 3-tray
configuration
(turning 90° the
tray support)
Porous material in
single package
1,00 0,30 0,25
Porous material in
double package
0,75 0,25 0,20
Solid and hollow
material in single
package
4,00 1,25 0,25
Solid and hollow
instruments in double
package
2,00 0,60 0,25
134 °C PRION 134 2,10 >18 B F L 53÷56 700 0,9
Porous, unpackaged
material
1,25 0,40 0,30
Porous material in
single package
1,00 0,30 0,25
Porous material in
double package
0,75 0,25 0,20
Hollow instruments in
single package
4,00 1,25 0,25
Solid and hollow
instruments in double
package
2,00 0,60 0,25
121 °C B CYCLE 121 1,10 20 B F L 54÷57 700 0,8
Porous, unpackaged
material
1,25 0,40 0,30
Porous material in
single package
1,00 0,30 0,25
Porous material in
double package
0,75 0,25 0,20
Hollow instruments in
single package
4,00 1,25 0,25
Solid and hollow
instruments in double
package
2,00 0,60 0,25
134 °C HOLLOW 134 2,10 4 S F C 34÷36 625 0,7
Unpackaged hollow
instruments
7,50 1,50 0,50
121 °C HOLLOW 121 1,10 20 S F C 48÷50 700 0,7
Unpackaged hollow
instruments
7,50 1,50 0,50
134 °C WRAPPED 134 2,10 4 S S L 31÷33 375 0,6
Solid instruments in
single package
4,00 1,25 0,25
We recommend
using the 3-tray
configuration
(turning 90° the
tray support)
121 °C WRAPPED 121 1,10 20 S S L 46÷47 400 0,6
Solid instruments in
single package
4,00 1,25 0,25
134 °C SOLID 134 2,10 4 N S C 24÷27 375 0,5
Unpackaged solid
instruments
7,50 1,50 0,50
121 °C SOLID 121 1,10 20 N S C 40÷42 400 0,5
Unpackaged solid
instruments
7,50 1,50 0,50
134 °C EMERGENCY 134 2,10 3 N S
Fast
14 375 0,45
Unpackaged solid
instruments
0,50 0,50 0,50
XXX°C CUSTOM
(see note)
134
or
121
2.10
or
1.10
> 4
or
> 20
n.d. F/S L/S n.d. n.d. n.d.
Unpackaged solid
instruments
n.d. n.d. n.d.
Variable
parameters
depending on
the settings
made
HELIX/BD TEST 134 2,10 3,5 - F C 22 - -
Test device only
(no other load)
- - -
VACUUM TEST - -0,80 - - - - 25 - - Empty chamber - - -