Specifications
APPENDIX A – TECHNICAL CHARACTERISTICS
55
APPENDIX A – TECHNICAL CHARACTERISTICS
SUMMARY TABLE
Device Steam Sterilizer
Classification (according to the Directive 93/42/EEC
and subsequent changes)
II b
Model
Manufacturer
M.O.COM. S.r.l.
Via delle Azalee, 1
20090 BUCCINASCO (MI) - ITALIA
Power supply voltage
220V - 240 V~
Frequency 50/60 Hz
Mains fuses
(6.3 x 32 mm)
F 16A 250V
On-board fuses
(5 x 20 mm)
F1: T 6,3A 250V (trafo secondary winding)
F2: T 3.15A 250V (trafo primary winding)
F1 PTR: T 3.15A 250V (printer protection)
External dimensions (HxWxD) (excluding rear
connections)
420 x 480x 560 mm
420 x 480x 660
mm
Nominal power 2300 W (10A)
Insulation class
Class I
Installation category
Cat. II
Environment of use
Internal use
Sound power level (A weighted) < 65 db(A)
Environmental operating conditions
Temperature: +15 °C ÷ +40 °C
Relative humidity: max 80%, non-condensing
Altitude: max 3000 m (a.s.l.)
Net weight:
empty
about 53 kg about 55 kg about 60 kg
empty with trays and support
about 54 kg about 57 kg about 62 kg
empty, with trays and supports and water at MAX
level
about 58 kg about 61 kg about 66 kg
Sterilization chamber dimensions
(Ø x D)
250 x 350 mm 250 x 450 mm
Sterilization chamber total volume about 17 l (0.017 m
3
)
about 22 l (0.022
m
3
)
Sterilization chamber useful volume
(with tray supports inserted)
about 10 l (0.010 m
3
)
about 13 l(0.013
m
3
)
Distilled water tank capacity
(supply)
about 4.6 l (water at MAX level)
about 0.8 l (water at MIN level)
Sterilization programs
Available: 11 (see Appendix B)
Pre-sets: 4 (direct selection by user)
Test programs
HELIX/BD Test
Vacuum Test
Preheating time
(from cold)
about 10 minutes
USB connection Standard female connector
Bacteriological filter
(PTFE filtering element)
Porosity: 0.2 μm
Connection: male 1/8" NPT connector