Specifications
11. TEST PROGRAMS
53
Close the door and start the program with the START key.
NOTE
I
F A PASSWORD HAS BEEN SET WITH THE
A
NY CYCLE START OPTION (SEE THE
CHAPTER, CONFIGURATION, SETTING THE PASSWORD), YOU WILL BE ASKED TO ENTER
THE ACCESS CODE
.
IN ADDITION, THE EQUIPMENT CHECKS THE PRINTER PAPER PRESENCE (OPTIONAL).
THE POSSIBLE WARNING MESSAGES, AND THE CONSEQUENT ACTIONS TO CARRY OUT,
ARE THE SAME AS DESCRIBED FOR A STANDARD STERILIZATION CYCLE
.
The vacuum phase begins immediately
and the display reads:
VACUUM TEST
VACUUMPULSE
‐ 0.69 bar01:02
The display shows the pressure (bar), and the total time from the start of the program.
When the pre-set pressure is reached (-0.80 bar) the pump stops and the pressure
stabilization phase begins (WAITING PERIOD), which lasts 5 minutes
(shown on the display
as a scalar value):
VACUUM TEST
WAITINGPERIOD
05:00
‐ 0.80 bar01:48
During this phase, a variation of the maximum low pressure is allowed of not more than 10%
,
without this causing the test to fail.
When the wait phase ends, the pressure verification
phase, true and proper, begins
(LEAKAGE PERIOD), with a duration of 10 minutes:
VACUUM TEST
LEAKAGEPERIOD
10:00
‐ 0.79 bar06:48
In this phase, a variation of up to ±0.02 bar
is allowed, compared to the initial phase value.
Higher variations cause the test to fail.
When this phase is also completed, the pressure is brought back to atmospheric pressure.
VACUUM TEST
LEVELLING
‐ 0.29 bar17:19