Specifications

9. RUNNING THE PROGRAM
49
The following message is then displayed:
MANUAL STOP
OPEN THE DOOR
85.8°CE999
0.01 bar26:01
Finally, when the door is opened, you will be asked to reset
the device by the following
message:
MANUAL STOP
RESETSYSTEM
85.5°CE999
0.01 bar26:01
To RESET the system, hold down, for at least three seconds, the PROGRAM SELECTION
key until you hear the confirming beep.
When the door is opened, the report for the sterilization cycle executed is produced, including
the error code (E999). Check the report, initial it in the space provided and file it in a suitable
place.
Refer to the print report examples shown in Appendix B
, Programs.
After the RESET, the device goes to STAND-BY mode, ready to execute a new program.
NOTE
W
HENEVER AN ALARM IS GENERATED IN CERTAIN PHASES OF THE CYCLE, AN
AUTOMATIC PROCEDURE IS ACTIVATED TO CLEAN THE PLUMBING CIRCUIT
. FOR A
COMPLETE DESCRIPTION OF THE ALARMS
, SEE APPENDIX E “ALARMS”.
WARNING
AFTER A PROGRAM IS M
A
NUALLY INTERRUPTED (MANUAL STOP)
ALWAYS CHECK THE STATUS OF THE ICON
BEFORE USING THE
MATERIAL IN THE STERILIZATION CHAMBER.
IF THE ICON
IS STEADY ON, THE MATERIAL IN THE STERILIZATION
CAN BE CONSIDERED STERILE AND, THUS, BE USED. WE RECOMMEND
USING IT IMMEDIATELY.
HOWEVER, IF IT IS OFF, THE MATERIAL IN THE STERILIZATION
CHAMBER CANNOT BE CONSIDERED STERILE AND ABSOLUTELY MUST
NOT BE USED.
> 3s