Specifications
 9. RUNNING THE PROGRAM 
  47 
Open the door 
Report print 
Equipment ready 
NOTE
A
T THE END OF THE CYCLE, AND UP TO THE OPENING OF THE DOOR, THE HEATING 
ELEMENTS ARE OFF
. AS A CONSEQUENCE, THE DEVICE IS SLOWLY COOLING 
REGARDLESS OF WHAT THE 
STAND-BY MODE IS. 
NOTE
W
HENEVER THE STERILIZER'S' DOOR IS NOT OPENED AT THE END OF THE CYCLE, THE 
VACUUM PUMP IS PERIODICALLY
 ACTIVATED TO REMOVE ANY TRACES OF CONDENSATE 
FROM THE STERILIZATION CHAMBER
. THE DISPLAY SHOWS: 
Press ⇑ to interrupt ventilation and open the door. 
Open 
the door and retrieve the sterilized material, using the extractor provided. 
The icon   symbol goes off. 
When the door is opened, the device goes to STAND-BY mode as previously set.. 
When the door is opened, the report for the sterilization cycle executed is automatically 
produced. Check the document, initial it in the space provided and file it in a suitable place. 
Refer to the print report examples shown in Appendix B, Programs. 
NOTE
I
F SELECTED THE PRINTOUT STEP BY STEP OPTION, THE REPORT WILL BE PRINTED 
DURING THE PHASES OF THE CYCLE
. 
NOTE
W
HEN A USB KEY IS INSERTED, IT IS ALWAYS POSSIBLE TO ELECTRONICALLY BACKUP 
THE PRINTING REPORTS
. 
The device is ready
 to execute a new cycle. 
Repeat the procedures explained in the Chapter, “Program Selection” for executing a new 
sterilization cycle. 
RESULT OF THE 
CYCLE 
After the cycle is finished, it is important to check the sterilization results. 
Whenever a cycle finishes (message CYCLE COMPLETE and icon 
on ), without, 
therefore, being interrupted by any type of alarm, you are guaranteed to have completely 
aseptic material. 
The report of the sterilization parameters is an additional check tool (and/or the check
madetotheparameterss avedonaUSBpendrive
). 
FORCED VENTILATION
⇑  to stop
35.2 °C
‐ 0.02 bar29:40