Specifications
6. CONFIGURATION 
34   
NOTE
I
F THE LAST CYCLE COMPLETED CORRECTLY (OR WAS INTERRUPTED BY MANUAL 
STOP) IT WILL BE POSSIBLE TO REPRINT IT IN EITHER NORMAL OR EXTENDED 
MODE
.  
I
F THE LAST CYCLE WAS INTERRUPTED BY AN ALARM (MANUAL STOP EXCLUDED) IT 
ONLY THE 
EXTENDED MODE WILL BE AVAILABLE. 
Following the reprint command, this message will be displayed: 
NOW PRINTING
THEREPORT
PLEASE WAIT...
which will remain on the screen until printing is finished. 
Setting the tank filling 
mode 
(FILLING OPTIONS on 
the ADVANCED menu) 
The internal tank can be filled either manually or automatically, in the latter case, drawing 
water from an external device (tank or demineralizer Milldrop connected to the device- see 
Chapter, “Installation”). 
After FILL OPTIONS is selected, the following menu appears: 
→
 AUTOMAT. FILLING + 
MANUAL FILLING↑ 
EXIT↓ 
‐
When AUTOMATIC FILL is selected, automatic filling is activated. 
In this case, as reached the minimum water level (icon MIN on) in the internal tank, the 
equipment enable the auxiliary water feeding pump for a pre-set time or a time needed to 
reach the maximum level (icon MAX on). 
When the maximum level (MAX signal) is reached, the automatic system is deactivated. 
NOTE
O
NLY ACTIVATE THE AUTOMATIC FILLING MODE 
A
FTER THE EXTERNAL TANK HAS BEEN 
FILLED WITH HIGH QUALITY DISTILLED WATER
 OR THE MILLDROP HAS BEEN TURNED ON. 
ALSO REMEMBER TO OPEN THE TAP ON THE EXTERNAL TANK OR THE MILLDROP. 
When MANUAL FILL is selected, the internal tank must be filled manually (see the Chapter, 
“First Start-Up”). 
Scroll through the items with the + and -  keys; confirm with the ↵ key. 
Setting the 
water draining mode 
(DRAIN OPTIONS from 
the ADVANCED menu) 
The water used for the sterilization cycle can be drained into either the internal
 tank (standard 
configuration) or an external
 tank of greater capacity (offered as an option – see chapter 
“Installation”) so as to reduce the frequency of emptying the used water. 
After DRAIN OPTIONS is selected, the following menu appears: 
→
 INTERNAL DRAIN+ 
EXTERNAL DRAIN↑ 
EXIT↓ 
‐