Specifications
6. CONFIGURATION
27
NOTE
T
O CHANGE THE PASSWORD, FIRST SELECT THE DISABLE OPTION, WHICH CANCELS
THE PREVIOUS PASSWORD
, AND THEN SELECT THE ANY POWER-ON OR ANY
CYCLE START OPTION, ENTERING THE NEW PASSWORD AS DESCRIBED ABOVE.
Setting the sterilization
programs
(PROGRAMS on the
ADVANCED menu)
The program setting and their storing in four pre-set positions is achieved in various steps
using several menus in sequence.
Each pre-set position can be associated to a standard or user configurable cycle (CUSTOM).
Let's look at the two cases separately.
To associate a standard program and define several of its parameters, proceed as follows:
1. Select PROGRAMS using the ↵ key; the following menu appears:
→
1st PRESET+
2nd PRESET↑
3rd PRESET↓
4th PRESET‐
EXIT
Define the position (1, 2, 3 or 4) to which the sterilization program will be associated using
the + and - keys. Confirm with the ↵ key.
2. From here, you enter the list of available cycles:
→
134 °C HOLLOW+
121 °C HOLLOW↑
134 °C SOLID↓
121 °C SOLID‐
134 °C EMERGENCY
134 °C WRAPPED
121 °C WRAPPED
134 °C B CYCLE
121 °C B CYCLE
134 °C PRION
XXX °C CUSTOM
EXIT
Using the + and - keys, scroll the list until you identify the sterilization program desired.
3. Confirm the selection with the ↵ key.
When the PRION program is selected, you will go to a screen for selecting the sterilization time.
TIME: XX min
+/‐to set
↵ to enter
⇑ to exit
A value can be set, starting from 18 minutes.