Specifications

6. CONFIGURATION
25
DEFAULTS
SETTINGS
The sterilizer leaves the factory with the following settings:
DATE: current date
TIME: current time
PROGRAMS: Preset 1: 134°C B CYCLE (standard drying)
Preset 2: 134°C HOLLOW (standard drying)
Preset 3: 134°C SOLID (standard drying)
Preset 4: 134°C EMERGENCY
NOTE
T
HE PROGRAMS INDICATED SHOULD BE CONSIDERED AS PREFERENTIAL SETTINGS.
HOWEVER, OTHER COMBINATIONS ARE POSSIBLE BASED ON THE DESTINATION MARKET.
ST-BY MODE: HIGH (preheating)
PRINT OPTIONS: (INTERNAL 1 copy, with optional printer
FILLING OPTIONS: MANUAL
DRAIN OPTIONS: INTERNAL
ACTIVATING
CONFIGURATION
OPTIONS
Setting the language
(LANGUAGE on the
BASIC Menu)
Now, we provide a detailed explanation of how to select the various available options,
proceeding in the shown in the previous paragraph.
Select LANGUAGE using the key. The following screen will appear:
ITALIANO+
ENGLISH
FRANÇAIS
DEUTSCH‐
ESPAÑOL
Select the desired language. Move using the + or keys and confirm using the key to store
the selection. After the data is confirmed, you return to the second-level menu.
NOTE
A
S SOON AS THE SELECTION IS CONFIRMED, ALL THE MENUS OF THE SETUP PROGRAM
WILL BE DISPLAYED IN THE LANGUAGE SET
.
Setting the date
(DATE SETTING on the
BASIC Menu)
When DATE SETTING is selected with the key, you will see:
dd/mm/yyyy
+/‐to set
to enter
to exit
Proceed as follows:
– The day flashes: set the current date with the + and - keys. Confirm with .
– The month flashes: set the current month with the + and - keys. Confirm with .
– The year flashes: set the current year with the + and - keys. Confirm with .
The date is stored. Once the last confirmation is given, you return to the second-level menu.