Specifications
4. INSTALLATION
16
CONNECTING
DEMINERALIZER
MILLDROP
The sterilizer can be connected to MILLDROP (water treatment system by reverse osmosis)
warranting the automatic reservoir filling with high quality demineralized water.
Refer to MILLDROP operating manual for the installation instructions.
NOTE
F
OR THIS OPTION SETTING, REFER TO CHAPTER “CONFIGURING THE DEVICE
–
AUTOMATIC FILLING OPTION”.
For additional information and advice about the correct connection of the sterilizer to the
various water purification systems, contact M.O.COM. customer support (see Appendix Z).
NOTE
B
ACKFLOW FROM THE MACHINE TO THE WATER CIRCUIT MUST BE PREVENTED BY USING
TOOLS THAT ARE IN CONFORMITY WITH LAW
IEC 61770.
DIRECT
CONNECTION TO A
CENTRALIZED
DRAINING POINT
Follow the instructions shown below for a correct direct connection to a centralized draining
point:
– Insert the silicone tube (provided) or other suitable plastic tube on hose union A; push the
tube all the way on and lock with the plastic tie or other means;
– Cut the tube to measure, push the free end on the connection provided on the centralized
draining point and lock with the plastic tie or other means;
NOTE
M
AKE SURE THE TUBE IS NOT BENT, CRUSHED OR OBSTRUCTED IN ANY WAY.
The following diagram provides an indicative arrangement of the components:
NOTE
T
HE CONNECTION POINT TO THE CENTRAL DRAIN MUST BE LOWER THAN THE
STERILIZER
'S SUPPORT SURFACE. OTHERWISE, THE TANK MAY NOT EMPTY CORRECTLY.
NOTE
F
OR THIS OPTION SETTING, REFER TO CHAPTER “CONFIGURING THE DEVICE
–
SETTING
THE WATER DRAINING MODE
”.
Clip
Support
plane
Centralized
draining point
Clip
Washer
Pipe This point must be at level
lower than the sterilizer's
support plane
To the centralized
draining point
A