Specifications
4. INSTALLATION 
10   
INSTALLATION 
INTRODUCTION 
The first and fundamental step in achieving good sterilizer operation, long life and complete 
use of its features is a correct, careful installation. Moreover, this precaution will avoid the 
danger of physical injury or property damage, not to mention malfunctions and damage to the 
machine. So, please follow the instructions in this chapter scrupulously. 
NOTE
M.O.COM.
 CUSTOMER SUPPORT (SEE 
A
PPENDIX Z) WILL ANSWER YOUR QUESTIONS 
AND PROVIDE ADDITIONAL INFORMATION
.  
T
HE STERILIZER HAS PASSED ALL REQUIRED INSPECTIONS BEFORE BEING PLACED ON 
THE MARKET
. IT DOES NOT REQUIRE ANY ADDITIONAL CALIBRATION BEFORE BEING 
PLACED IN SERVICE
. 
Dimensions and weight
B and B+  B
2
A. Height (total)  420 mm  420 mm 
B. Width (total)  480 mm  480 mm 
C. Depth (excluding rear 
connections) 
560 mm  660 mm 
Total weight  58 kg  63 kg 
Electricity 
The electrical system to which the sterilizer will be connected must be suitably dimensioned 
based on the electrical characteristics of the device. This information is shown on the back of 
the machine. 
COMPARTMENT 
DIMENSIONS FOR 
BUILT-IN 
INSTALLATIONS 
When installing the sterilizer inside a cabinet, you must 
provide adequate space all around the device to 
provide effective ventilation as well as a large enough 
opening in the back that, in addition to allowing the 
passage of the power cord will also provide an 
adequate air flow and the consequent optimum cooling 
of the heat exchanger. 
The compartment where the steriliser will be kept must 
have the following minimum dimensions: 
Dimensions and weight
B and B+  B
2
A. Height  500 mm  500 mm 
B. Width  580 mm  580 mm 
C. Depth  600 mm  700 mm 
WARNING
COMPARTMENT DIMENSIONS LESS THAN THOSE SHOWN MAY 
COMPROMISE THE CORRECT CIRCULATION OF AIR AROUND THE 
DEVICE AND MAY NOT PROVIDE ADEQUATE COOLING, WITH THE 
CONSEQUENT DETERIORATION OF PERFORMANCE AND/OR POSSIBLE 
DAMAGE. 
NOTE
I
F THE MAIN SWITCH IS INACCESSIBLE WHEN INSTALLED IN THE COMPARTMENT, USE AN 
ELECTRIC PLUG THAT INCORPORATES AN ON
/OFF SWITCH. 
D
O NOT REMOVE THE UPPER COVER OR ANY OTHER EXTERNAL PART. WHEN INSTALLED 
IN THE COMPARTMENT
, THE DEVICE MUST BE COMPLETE WITH ALL ITS PARTS. 
P
LEASE REFER TO APPENDIX A (TECHNICAL CHARACTERISTICS) FOR COMPLETE 
TECHNICAL DATA
. 