Cluster Configuration: Within a cluster, one of the
cabinets must be a control cabinet. The remaining
cabinets in the cluster are auxiliary cabinets. The
maximum number of auxiliary cabinets that can be
placed on a single side of a control cabinet is 3.
The maximum number of cabinets (including the
control cabinet) in a cluster is 5.
Door Swing: Cabinets can be ordered with doors
that open from the right or left. The door swing
needs to be identified at the time the order is
placed. The above figure shows doors that open
from the right (left hinge.)
Figure 40 - iRISupply™ 'Patient List' Screen
3. Select a patient on the list, or click on the
‘Override’ button. This will take you to the ‘cluster’
security configuration screen (Figure 41). This
screen will indicate when doors in the cluster you
will have access to. Validate that the doors
highlighted in green are unlocked and the doors
highlighted in red are locked.
Cabinet Access: It will be necessary to have access
to the rear of the cabinets for service and
maintenance. Because the cabinets are on wheels,
they may be placed against the rear wall during
normal operation. The site must have enough
available space to roll the cluster out from the wall
during service calls.
Figure 41 - iRISupply™ Cluster Security Screen
4. Open a door and remove items from its
compartment, and then close the door. The door in
question will change color from green to red, and
4