Installation Guide
Document Name: iRISupply™ Installation and Assembly Guide
Published: 05/11/2004 Doc #: Version 2.0
Figure 20 - Electrical Connections in the Powered Auxiliary
Cabinet
3.2. Mechanical Assembly
The mechanical assembly of iRISupply™ involves
rolling the cabinets into place, bolting them together
and leveling the cabinets.
3.2.1. Securing the Service Panels
When the physical and wiring connections for the
cabinets have been completed, you must securely
enclose the iRISupply™ System by mounting all of the
required service panels. Each iRISupply™ Cabinet
(including both the master and auxillary units), contain
four service panels - 2 Side Panels, 1 Top Panel, and 1
Back Panel (Figure 21).
Side Panel
Side Panel
Top Panel
Back Panel
iRISupply
TM
Front View iRISupply
TM
Rear View
Document Name: iRISupply™ Installation and Assembly Guide
Published: 05/11/2004 Doc #: Version 2.0
19
3.1.4. Antenna Tuning
Each matching box must be tuned prior to system
power up. Complete the following steps to tune the
matching box.
• Remove the cover from the matching box (Figure
15).
Figure 15 - Matching Box
• Disconnect the antenna cable. Connect one end of
a 24 ft co-axial cable to the matching board, and
the other end to the SWR Meter (Figure 16).
Ensure that the SWR Meter is set to 13.56 MHz.
Figure 16 - Matching Board to SWR Meter Connection
• Utilizing a flat-head screwdriver, adjust the three
variable capacitors (Series 1, Parallel, and Series 2
in Figure 17). First, adjust the Parallel variable
Figure 21 - iRISupply(TM) Cutaway with Service Panels Visible
22