Installation Guide

Document Name: iRISupply™ Installation and Assembly Guide
Published: 05/11/2004 Doc #: Version 2.0
Table 3 – Number Antenna Cable by
Cabinet Type
Type of Cabinet Number of antenna
cables
Standard 3
Standard w/Power 3
Control, Standard 2
Control, Stent 3
Catheter 1
Dual-Catheter 2
Stent 5
Each antenna cable will be numbered. Cable number
one will always be the top antenna in the cabinet. The
highest numbered cable will be the bottom antenna.
The antenna connectivity cables will be routed through
the wire tray on the top of the cabinets and into the
control cabinet. In the control cabinet, the antenna
connectivity cables will be connected to the appropriate
port on the multiplexer. The multiplexer has 19 BNC
connectors. The connectors serve the following
functions:
Table 4 – Connector Function on Main
Multiplexer
Number Function
1 Dummy Load
2 through 8 Antenna Connections
9 Output to Additional MUX #1
10 RFID Reader Input
Connection
11 Output to Additional MUX #2
12 through 19 Antenna Connections
Document Name: iRISupply™ Installation and Assembly Guide
Published: 05/11/2004 Doc #: Version 2.0
25
Level each cabinet so that they are level and at the
same height as its neighboring cabinet. This is
most easily done by rolling the cabinets around on
the floor until they are at the same height. The
leveling legs should be utilized to assist in the
leveling effort.
There are 4 bolts that hold two cabinets together.
Two are in the front (top and bottom) and two are
in the rear (top and bottom.)
The cabinets are bolted together using the lugs at
the top and bottom of the cabinets. Use a ½”
wrench to tighten the bolts. A flat washer should
be used on the bolt and the nut. Figure 25 shows
how the front bolts are installed.
Figure 25 - Cabinet Bolts
NOTE
Prior to bolting the cabinets together, remove the
back panels to verify that the antenna cables and
matching board power cables are properly
connected to the matching boards.
3.2.4. Connecting Power to iRISupply™
When the physical and wiring connections have been
completed, connect iRISupply™ to a standard power
outlet.
At the rear of the iRISupply™ Master Cabinet is a
power socket connector (Figure 26).
16