User manual
PASS-FAIL MODE
65
6565
65
Frequency (except
Rdc)
Press the Left/Right key repeatedly to
move the cursor to Frequency, and use
the numerical and unit keys to enter the
value.
2.00 195.00 kHzVac
Frequency step
resolution
For frequency increase/decrease using Up/Down
keys, fine and coarse step settings are available.
Press the Code key and enter 10 (Fine)
or 11 (Coarse).
Code
Voltage
Press the Left/Right key repeatedly to
move the cursor to Voltage, and use the
numerical and unit keys to enter the
value.
195.002.00 Vac kHz
LCR-8000G Series User Manual
66
6666
66
Single-Step Test Run
Run in Absolute mode
1. Press F5 to select Absolute
measurement.
F 5
2. Use the Left/Right key to move the
cursor to Hi/Lo value for editing.
HiLo 20.0mF0.00mF
3. Use the numerical keys
and unit keys to enter the
value. The editing value
appears at the bottom left
corner of the display.
Range
Speed Med
Auto
Lo Lim: 1.5_
Example
1.2m
m
Ω
Enter
Backspace
All clear
Clear
Increase
Decrease
Hi and Lo values are automatically swapped if
necessary
Hi and Lo Swapped
4. The display updates the Hi/Lo result
immediately. The result is pass if the bar stays
in the central area. The buzzer sounds
accordingly.