User manual

BASIC MEASUREMENT
57
5757
57
DC
AC≤
100Hz
AC≤
2kHz
AC>
2kHz
AC≥
1MHz
Slow
900ms
1.3s 600ms
600ms
620ms
Med
120ms
1.2s 470ms
450ms
470ms
Fast
60ms 650ms
180ms
150ms
150ms
Max
30ms 600ms
120ms
75ms 120ms
Beep setting
If the beep setting (page61) is active and
the display is in Pass/Fail test mode, it
might sound continuously depending
on the measurement result. If this
happens, press the Sing/Rep key and
set the mode to Single. Then turn Off
the buzzer.
Sing/Rep
Hide drive Voltage/Current
Background
The drive voltage and current shows
the actual voltage/current level
applied to the device under test.
Vm:549.8mV
Im:724.9pA
Panel operation
1. Press the Code key.
Code
2. Enter the system code using the numerical
keys, then press the Enter key.
Hide drive V/I: 80
Enter
Vm:549.8mV
Im:724.9pA
Show drive V/I: 81
Enter
Vm:549.8mV
Im:724.9pA
LCR-8000G Series User Manual
58
5858
58
PASS-FAIL MODE
In the Pass/Fail test mode, measurement results are
compared with user-defined limits and the results
are displayed. Two types of tests are available: Single
and Multi-Step. The Single test shares the same
interface with the basic measurement, and tracks one
item. The Multi-Step test runs a program comprised
of multiple measurement items with different
parameters.
Single-step
configuration
Overview ......................................................................60
Configure beep setting ................................................61
Configure the Average .................................................62
Select test item and scale (Pass/Fail test)..................63
Set parameters.............................................................64
Single-step run
Run in Absolute mode.................................................66
Run in Percentage mode .............................................67
Run in Delta mode.......................................................68
Use display value as nominal......................................69