User`s manual

3
INP-TN5A0202-E
In cases where damage or problems with this equipment
may lead to serious accidents, install appropriate external
protective circuits.
To prevent damage and failure of the equipment, provide
the rated power voltage.
To prevent electric shock and equipment failure, do not
turn the power ON until all wiring is complete.
Before turning the power ON, confirm that clearance
space has been secured to prevent shock or re.
Do not touch the terminal while the machine is ON. Doing
so risks shock or equipment errors.
Never disassemble, convert, modify or repair this
equipment.Doing so risks abnormal operation, shock or
re.
All of the wiring should be class 1 type wiring or the low
voltage wires are routed separately from the hazardous
voltage wires to ensure separation of circuits.
When using a AWG-16 cable, you should use the crimp
terminal that material thickness is 0.9mm or less.
Maintenance
When installing or removing the equipment, turn the power
OFF. Otherwise, shock, operational errors or failures may
be caused.
Periodic maintenance is recommended for continuous and
safe use of this equipment.
Some parts installed on this equipment have a limited life
and/or may deteriorate with age.
The warranty period for this unit (including accessories) is
one year, if the product is used properly.
Caution
Cautions when Installing
For install in UL listed enclosure only.
Please avoid installing in the following locations.
Locations in which the ambient temperature falls outside
the range of 0 to 50 degrees C when equipment is in use.
Locations in which the ambient humidity falls outside the
range of 45 to 85% RH when equipment is in use.
Locations with rapid temperature changes, leading to dew
condensation.
Locations with corrosive gases (especially sulfide gas,
ammonia, etc.) or ammable gases.
Locations with vibration or shock directly.
Please Read First (Safety Warnings)
Please read this section thoroughly before using and
observe the mentioned safety warnings fully.
Safety warnings are categorized as "Warning", "Caution" or
"Risk of Electrical Shock".
Warning
Improper use of the equipment may result in death or
serious injuries.
Caution
Improper use of the equipment may cause injury or
property damage.
Risk of
Electrical
Shock
Indicates that a risk of electrical shock is present and
the associated warning should be observed.
Warning
Installation and Wiring
This equipment is intended to be used under the following
conditions.
Ambient temperature
-
10 to 50 degree C
Ambient humidity 90% RH or below (with no condensa-
tion)
Vibration 10 to 70Hz less than 9.8m/s² (1G)
Warm-up time 30 min. or more
Installation category IEC1010-1: class II
Pollution level IEC1010-1: degree 2
Between the temperature sensor and the location where
the voltage reaches or generates the values described
below, secure clearance space and creepage distance as
shown in the table below.
If such space cannot be secured, the EN61010 safety
compliance may become invalid.
Voltage used or generated by
any assemblies
Clearance
Space [mm]
Creepage Space
[mm]
Up to 50 Vrms or Vdc 0.2 1.2
Up to 100 Vrms or Vdc 0.2 1.4
Up to 150 Vrms or Vdc 0.5 1.6
Up to 300 Vrms or Vdc 1.5 3.0
Above 300 Vrms or Vdc Please consult our distributor
For the above, if voltage exceeds 50Vdc (called danger
voltage), functional insulation is required between the
earth and all terminals of the equipment.
Note that the insulation class for this equipment is as
follows. Before installing, please con rm that the insulation
class for equipment meets usage requirements.
RS-232C communication
RS-422 communication
RS-485 communication
Functional insulation (AC500V) Functional insulation (AC1000V)
Power supply