User`s manual
3 - 1
MELSEC-
A
3 SPECIFICATION
3 SPECIFICATION
This chapter provides the specifications of the AJ65SBT-64AD.
3.1 General Specification
Table 3.1 indicates the general specifications of the AJ65SBT-64AD.
Table 3.1 General specification
Item Specification
Usage ambient temperature 0 to 55°C
Storage ambient temperature -20 to 75°C
Usage ambient humidity
10 to 90% RH, non-condensing
Storage ambient humidity
Vibration durability
Compliant with
JIS B 3502
and IEC
61131-2
Frequency Acceleration Amplitude Sweep count
Under intermittent
vibration
5 to 8.4Hz —
3.5mm
(0.14 inch)
10 times each
in X, Y, Z
directions
8.4 to 150Hz 9.8m/s
2
—
Under continuous
vibration
5 to 8.4Hz —
1.75mm
(0.069 inch)
—
8.4 to 150Hz 4.9m/s
2
—
Shock durability
Compliant with JIS B 3502 and IEC 61131-2
(147 m/s
2
, 3 times each in 3 directions X, Y, Z)
Operating atmosphere No corrosive gases
Operating altitude
*3
0 to 2000m
Installation location Inside a control panel
Overvoltage category
*1
Less than II
Pollution level
*2
Less than 2
*1 This indicates the section of the power supply to which the equipment is assumed to be connected between the
public electrical power distribution network and the machinery within premises.
Category II applies to equipment for which electrical power is supplied from fixed facilities.
The surge voltage withstand level for up to the rated voltage of 300V is 2500V.
*2 This index indicates the degree to which conductive material is generated in terms of the environment in which the
equipment is used.
Pollution level 2 is when only non-conductive pollution occurs. A temporary conductivity caused by condensing must
be expected occasionally.
*3 Do not use or store the programmable controller under pressure higher than the atmospheric pressure of altitude 0m.
Doing so may cause malfunction. When using the programmable controller under pressure, please consult your
local Mitsubishi representative.