User`s manual
-9-
3. Handling Instructions   
A.  Sterilization of Endoscope Holder 
Warning:  The Point Setter arm and OR table adapter are never sterilized. Instead 
 they are covered by a sterile drape as referenced in 3. D. (page 15). 
  Sterilize the endoscope holder using the following instructions. Please follow these 
instructions carefully. 
CLEANING AND STERILIZATION OF REUSABLE ACCESSORIES 
FOR THE MITAKA POINT SETTER (“ENDOSCOPE HOLDER”) 
NOTE: Throughout this sterilization section, the reusable sterilizable accessories that 
are attached to the Point Setter arm through the sterile drape, as shown on page 
10, are collectively referred to as “endoscope holder” for purposes of 
simplification. 
Warning: The endoscope holder must be thoroughly cleaned and sterilized according to 
validated infection control procedures prior to use and subsequent reuse. 
Caution:   Any deviations from the recommended parameters for cleaning and sterilization 
should be validated by the user. 
■ Preparation for Cleaning and Sterilization
Place the endoscope holder in a container and soak with a neutral pH (pH 
5.0 to 8.0) enzymatic cleaning solution (e.g., Enzol, Metrizyme, or 
equivalent per manufacturer’s instructions) immediately after use to 
prevent blood, protein and other contaminants from drying onto the 
endoscope holder. 
■ Water Quality Requirements 
Distilled water is recommended for cleaning and rinsing of the endoscope 
holder. 