User`s manual

-19-
4. Further Safety Instructions
1. The apparatus must be operated only by an expert. (USA requires a physician approval.)
2. In installing the apparatus pay attention to the following matters:
(1) Do not install on a wet place and be sure not to splash water or other liquid.
(2) Do not use this apparatus outside the operating room.
(3) Great care is necessary in handling the apparatus to avoid a dent or a scrape.
(4) Be sure to firmly tighten all clamps and brakes.
(5) The maximum permissible weight must not exceed 2kg.
(6) Do not use parts which are not products of MITAKA KOHKI Co. Ltd.
3. Pay careful attention to the following matters before use:
(1) Be sure to perform operational check before operation.
(2) Handling of the apparatus requires great care. In case of knocking or hitting the apparatus in use, be
sure to stop using it and contact MITAKA KOHKI Co. Ltd.
(3) Make sure that each cable has no damage. In case of finding defects, be sure to stop using the
apparatus and contact MITAKA KOHKI Co. Ltd.
(4) Use nitrogen gas or pressurized air with designated operating pressure.
(5) Be sure to securely connect hoses of nitrogen gas or pressurized air.
4. Pay careful attention to the following matters in use:
(1) Care must be taken not to exceed time needed for diagnosis and treatment.
(2) Constant observation is required over the apparatus and the patient.
(3) The operating room temperature should be 15~29 degrees C or 59~84 degrees F, and humidity
45~90%, respectively.
5. Pay careful attention to the following matters after use:
(1) Do not disconnect hoses by pulling out with excessive force.
(2) The storage place requires the following care:
a. Store in a dry place. Do not splash with water or other liquid.
b. Do not store the apparatus in a place susceptible to vibration or shock.
c. The storage temperature should be room temperature (approximately 20 degrees C or 68
degrees F).
(3) Attachments and hoses should be stored together after cleaning.
(4) Do not use an organic solvent to clean the apparatus.
(5) If you recognize damage or deformation on the surface of the hose, stop using the apparatus and
contact MITAKA KOHKI for inspection.
6. Failure:
In the event of a failure, indicate the problem properly. An appointed MITAKA KOHKI Co. Ltd. technical
service person or those who are trained by MITAKA KOHKI Co. Ltd. must repair the apparatus.
7. Maintenance:
Be sure to perform an operational check before use and make sure the unit has no damage or malfunction.
In additional to an operational check each use Mitaka Kohki recommends that the Point Setter inspection
and operation test in Section 5 below be performed at least monthly.