User`s manual
-14-
3. Connect hoses
Nitrogen gas or compressed air pressure should be
6 ~ 8 kgf/cm or 90 ~ 115 psi or 6.2 to 7.9 bar.
Excessive
p
ressure is ver
y
dan
g
erous.
4. Operational check
Perform operational check of the
Point Setter and make sure the
range of motion is appropriate.
Do this before starting surgery
and follow Section 5 on a monthly
basis.
NOTE: If instrument does not
release completely or is slow to
release, then recheck nitrogen or
air pressure. Lower than
recommended pressure will cause
poor performance.
!
CAUTION!
Plug the arm hose into the
nitrogen connector of the hospital.
Set pressure to 6 ~ 8 kgf/cm or 90
~ 115 psi or 6.2 to 7.9 bar.
NOTE: The PSMS comes with a
male Schrader connector. Other
connectors or adapters can be
acquired.
2
Connector from
“Point Setter” arm
Connector of Nitrogen hose
(PSNS-060)
a male Shrader connector
(PSNS-060)
2