User`s manual

-13-
C. Before Draping Patient
Perform this part before covering the patient with sterile cloth.
Set up OR table adapter on the side rail of the operating table. Try to locate the arm
at the best position during operating considering the C-arm X-ray or other possible
interfering surgical instruments. We recommend that you confirm interference using real
equipment and finish this setup before beginning operating. Otherwise, changing the
position of the arm during the operation can be very difficult. To change the OR table
adapter during surgery a non-sterile person may be needed.
Set up Point Setter arm. Make sure that
the arm is securely clamped to the OR
table adapter.
Clamp the OR table adapter
firmly to the side rail. Shake
unit to insure that it is
secured firmly to the side
rail.
Disengaging the cap (right
and left) enables you to
replace each part. Select the
order according to the setup
positioning to the operating
table or operating method.
1. Setup OR table adapter to the operating table
2. Setting up the arm