User`s manual
-13-
C. Before Draping Patient  
  Perform this part before covering the patient with sterile cloth. 
  Set up OR table adapter on the side rail of the operating table. Try to locate the arm 
at the best position during operating considering the C-arm X-ray or other possible 
interfering surgical instruments. We recommend that you confirm interference using real 
equipment and finish this setup before beginning operating. Otherwise, changing the 
position of the arm during the operation can be very difficult. To change the OR table 
adapter during surgery a non-sterile person may be needed. 
Set up Point Setter arm.    Make sure that 
the arm is securely clamped to the OR 
table adapter. 
Clamp the OR table adapter 
firmly to the side rail. Shake 
unit to insure that it is 
secured firmly to the side 
rail. 
Disengaging the cap (right 
and left) enables you to 
replace each part. Select the 
order according to the setup 
positioning to the operating 
table or operating method. 
1. Setup OR table adapter to the operating table 
2. Setting up the arm 










