User Guide
11 • Miro Canvas • User Guide
On-screen Error: Voltage
Using the provided leveling plate, adjust the feet of
the instrument to ensure the bubble is in the center,
indicating a level state. If error persists, please contact
Miroculus Technical Support.
Voltage is not detected and run cannot proceed. Please
ensure cartridge clamp is engaged, press cancel, and
restart the run.
The lid has not been properly closed. Press down on
both front corners of the lid to ensure full closure. If error
persists, press cancel and restart the run.
Cartridge may have shifted or clamp may not be properly
closed. Open the instrument lid, release the cartridge
clip engages. Securely close the lid by pressing down
on both corners. If error persists, discard the current
cartridge and load a new cartridge.
RUNNING A PROTOCOL
On-screen Error: High Humidity
On-screen Warning or Error: High Temperature
The environment that the instrument is located in is too
cold. Ensure the environmental temperature is within the
operating range (19-25°C). Run will fail at 16°C.
The environment that the instrument is located in is too
humid. Ensure the environmental humidity is within the
operating range (20 - 80%, non-condensing).
The environment that the instrument is located in is too
hot. Ensure the environmental temperature is within the
operating range (19-25°C).
Miroculus Technical Support.
On-screen Error: Magnet Issue
On-screen Error: Voltage
On-screen Error: Heating System
cartridge and load a new cartridge.
and load a new, unexpired cartridge.
The instrument is not functioning as expected. Recover
your sample, if possible. Please restart the instrument
contact Miroculus Technical Support.
Voltage is not detected and run has failed. Please
contact Miroculus Technical Support.
The instrument is not functioning as expected. Run
has failed and an attempt will be made to recover your