4.2 Attaching the BLE 3000 module
Carefully insert 26-pin connector, ensuring side labeled “BATTERY SIDE” is inserted closest to the battery (5).
Check the cleanliness of the module gasket.
Engage the module bottom side first, with the gasket placed inside the bottom of the DMC 3000.
Rotate the module (7).
Hand tighten the two module screws with the provided screwdriver (1) at 0.25 to 0.35 Nm (8).
Ensure the icon is displayed. If it does not, verify the dosimeter is configured to accept the module (see
§2.2.3 and §2.7).
Module Attachment
Note:
Due to the presence of voltage on the pins of the 26-pin connector, remove or add new modules when the dosimeter
is in Pause mode. Avoid shorting the pins.
(5)
(6)
(8)