User's Guide BLE 3000 Module Radiation Safety. Perfected.
Table of revisions BLE 3000 Module User's Guide -2DOC015145EN-A Revision Date Verified by Approved by Origin and description of the revision Edited pages A 2020-07 T. VIALLAT V.
Table of Contents BLE 3000 Module User's Guide 1- Presentation........................................................................... 4 5- Maintenance........................................................................13 2- General.................................................................................. 5 5.1 Maintenance........................................................................... 13 2.1 Use and Functionality............................................................
1- Presentation BLE 3000 Module User's Guide Connection pad Attachment screws Wearing and Battery Guidelines Technical information The BLE 3000 Module is mounted on a DMC 3000 to expand the DMC 3000’s measurement capabilities to Bluetooth features. This user’s guide is to be used along with the DMC 3000 User’s Guide (n°151153EN).
2- General BLE 3000 Module User's Guide 2.1 Use and Functionality The BLE 3000 module is part of the accessory range of Mirion Technologies DMC 3000 electronic dosimeter. When attached to the DMC 3000, the BLE 3000 module adds the following features: Bluetooth functions, Amplification of visual alarms, Alarm report to an external accessory (wristband, smartwatch...). The BLE 3000 module physically integrates into the dosimeter's case. 2.1.
General BLE 3000 Module User's Guide 2.2 Settings The BLE 3000 module contains all the proper functional settings parameters. These parameters are saved in the module on non-volatile memory. 2.3 Bluetooth Functions 2.3.1 Transmission When the dosimeter switches to measurement mode, the BLE 3000 module automatically switches to transmission mode. The BLE 3000 module then periodically transmits the dosimetric data, the normal transmission rate can be set between 2 and 255 seconds.
General BLE 3000 Module User's Guide 2.4 Historical Data Function Measurements and events of the BLE 3000 module are automatically stored in the DMC 3000 histogram. 2.5 Handling Recommendations Do no use the dosimeter belt clip when the BLE 3000 module is attached as the opening gap of the clip is significantly reduced. The user must always wear the dosimeter against the body (see the label on the back of the module).
General BLE 3000 Module User's Guide 2.6 Module Management The Module Management Mode is configurable using DMCUser software. None: any module connection is ignored by the dosimeter. All: DMC 3000 accepts any type of modules. When a module is connected, the dosimeter automatically changes its display configuration and displays the Hp(10) neutron or Hp(0,07) shallow dose depending on the type of module.
3- Alarms BLE 3000 Module User's Guide 3.1 Bluetooth Function Fault and the DMC 3000 During a fault with the telemetry function of the BLE 3000 module, the DMC 3000 will exhibit the following characteristics: Dosimeter detection will not be affected by the telemetry function failure. Dosimeter will retain all Dose, Max Rate, set points and status bits regardless of module status, failure or connections. Dosimeter Dose and Dose Rate Alarms will take precedence over any telemetry fault messages.
4- Module Attachment BLE 3000 Module User's Guide (1) (3) - (2) (4) + 4.1 Removing the dosimeter battery cache Make sure the dosimeter is in PAUSE mode. Unscrew the two screws (2) with the screwdriver provided with your DMC 3000 (1). Rotate the battery cover (3). Remove the battery cover (4). Note : Due to the presence of voltage on the pins of the 26-pin connector, remove or add new modules when the dosimeter is in Pause mode to avoid shorting the pins.
Module Attachment BLE 3000 Module User's Guide 4.2 Attaching the BLE 3000 module Carefully insert 26-pin connector, ensuring side labeled “BATTERY SIDE” is inserted closest to the battery (5). Check the cleanliness of the module gasket. Engage the module bottom side first, with the gasket placed inside the bottom of the DMC 3000. Rotate the module (7). Hand tighten the two module screws with the provided screwdriver (1) at 0.25 to 0.35 Nm (8). Ensure the icon is displayed.
Module Attachment BLE 3000 Module User's Guide (8) (9) (10) (11) 4.3 Detaching the BLE 3000 module Make sure the DMC 3000 is in Pause mode. Unscrew the two screws (8) with the screwdriver provided with your DMC 3000 (1). Pivot the module and detach it from the dosimeter (9). Conduct a software disconnect using the "Detach Module" command of DMCUser software (10), or temporarily remove the dosimeter’s battery (11). Replace the battery if necessary.
5- Maintenance BLE 3000 Module User's Guide -13DOC015145EN-A 5.1 Maintenance Mirion Technologies recommends full inspections of dosimeters and modules at a minimum of once a year. Functions: check that the DMC 3000 displays .
6- Technical Characteristics BLE 3000 Module User's Guide -14DOC015145EN-A 6.1 EU declaration of conformity EU declaration of conformity for BLE 3000 module: DOC013812 The EU declaration of conformity certificate can be delivered on request by Mirion Technologies SAS. 6.2 Electrical Characteristics DMC 3000 battery life (with the BLE 3000 module) is 4.5 calendar months (typical use of eight (8) hours a day, five days per week in Run mode, without excessive alarms (*).
Technical Characteristics BLE 3000 Module User's Guide 6.3 Mechanical Characteristics Rugged, high impact polycarbonate-ABS case. Dimensions (DMC 3000 + BLE 3000 module): 60 x 88 x 29 mm (2.4 x 3.5 x 1.1 in) with standard clip 86 x 56 x 21 mm (3.4 x 2.2 x 0.8 in) without clip. Weight (BLE 3000 module + DMC 3000 with alkaline batteries and a standard clip): < 105 g (3.7 oz). Weight (BLE 3000 module only): 20 g (0.7 oz) 6.
Technical Characteristics BLE 3000 Module User's Guide 6.6 Environmental regulations In accordance with European regulation (Directive DEEE 2002-96-CE), BLE 3000 modules should not be thrown out with non-sorted waste, but must be selectively collected. Follow local regulations. 6.7 FCC regulations FCC ID: 2AWTM-BLE3000 This device complies with Part 15 of the FCC Rules.
Technical Characteristics BLE 3000 Module User's Guide 6.8 IC regulation IC number: 26297- BLE3000 This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada's licence-exempt RSS(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
6.
BLE 3000 Module User's Guide Notes: -19DOC015145EN-A
BLE 3000 module User's Guide www.mirion.