Manual

Health Physics
Division
Since norms, specications and designs are subject to occasional change, please ask for conrmation of the information given in this publication.
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
DESCRIPTION
1 detection unit including 1 NaI(Tl) detector (2 or 4 liters),
digital MCA
1 ruggedized tablet PC
1 GPS, WAAS and EGNOS compatible
SIA identication algorithm designed for challenging
Homeland Security and nuclear accident fallout issues
Simplied SpirMOBILE interface, based on Google
Earth or georeferenced views or maps
Data export to ESRI Geographical Information System
SpirPORTAL interface for deployable portal use
FUNCTIONS
0.5 or 1 second continuous elementary spectra acquisi-
tion and stabilization (sourceless)
Count rate/dose rate prole display
Continuous sliding spectra identication
Alarm ltering according customer criteria
Simplied panel display ( level/ ID; instant or trend)
Expert display ( spectra, trend, raw and ltered ID)
Mapping: trajectory with level indication and alerts
locations ags
Georeferenced database, replay and data export
capabilities
PERFORMANCES
Typical 4 liters detector: for
137
Cs, 8%
resolution and 37cps per nSv/h sensitivity
1024 channels fast MCA, one per detector
Energy range 25 keV to 3 MeV
Identication up to 4 isotopes mixed (in addition to
background), with condence level and quantication
indication
Isotopes list according to ANSI N42-34, N42-38,
IEC62327, IEC 62484 draft, IAEA Tecdoc
User selectable alarm criteria
CHARACTERISTICS
Polyester resin strengthened with glass ber containers
Size:
Single detector case: 84 cm x24 cm x24 cm (33 in x9
in x9 in), 18 kg (39.7 lb) (with 2 l NaI(TI)), 24 kg (52.9 lb)
(with 4 l NaI(TI))
Operating temperature range:
-20°C to +50°C (-68°F to +122°F), temperature shock
protected
Detector unit power supplied by the tablet PC
SPIR-IDENT Mobile light
Detection & Identication - Risks Prevention
www.mirion.com
149026EN-C
NORM:
40K, 226Ra and daughters, 232Th and
daughters
Medical:
18F, 51Cr, 67Ga, 99Mo , 99mTc, 103Pd,
111In, 123I, 125I, 131I, 133Xe, 153Sm, 201Tl
Industrial
and Nuclear
Accident:
22Na, 57Co, 60Co, 75Se, 106Rh, 132I, 133I,
133Ba, 134Cs, 137Cs, 152Eu, 192Ir, 241Am
SNM:
233U, 235U, 238U, 239Pu, 241Pu, 237Np