Owner's manual

Health Physics
Division
Since norms, specications and designs are subject to occasional change, please ask for conrmation of the information given in this publication.
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
SpiR-ID
Detection & Identication - Risks Prevention
NUCLEAR CHARACTERISTICS
Detectors
SpiR-ID NaI: 3” x 1.5”
SpiR-ID LaBr: 1.5” x 1.5” LaBr
2 moderated LiI(Eu) detectors
GM tube for high gamma dose rate
Energy range
25 keV to 3 MeV (gamma)
0,025 eV to 15MeV (neutrons)
Measurement range
<0.01 µSv/h to 10 mSv/h ( 1µR/hr to
1 R/hr)
Identification
fast digital MCA 1024 channels
throughput >100 000 cps
range <0.01 to >25 µSv/h (137Cs)
detection and ID performance
overpasses ANSI N42-34, IEC62327,
IAEA NSS1 recommendations
Continuous spectra acquisition and
stabilization
no need of eld calibration
Continuous count rate analysis
versus dynamic background
True dose rate calculation by spectra
weighting
Various spectra accumulation mode:
sliding / automated / manual
FUNCTIONAL FEATURES
Large TFT 3.5” VGA (640 x 480) sun
readable display
trend curve screen with level and ID
spectra and ID combined screen
event list screen with easy scroll
Alarm indication
sound or voice alarm and chirp
vibrator in the handle
colored LEDs
progressive alarm based on
identication and level
easy acknowledge
safety alarm
Routine mode / Expert mode
Embedded GPS
Data storage: SD card
Data retrieve and remote operation
USB link for PC host
External alpha/beta probe connection
IDENTPRO/SIA IDENTIFICATION
ALGORITHM
Designed for reliable decision with
short spectra
Solves mixed isotopes (up to 8)
Processes masking scenarios
Tolerant to shielding
Category, isotope, condence level
indication
ENVIRONMENTAL, ELECTRICAL,
MECHANICAL CHARACTERISTICS
Temperature range: –20°C to 50°C
(-4°F to 122°F); -40°C on request
Humidity: up to 100%
Shock, vibration and drop resistant
MIL461D EMI compliant
Water and dust ingress IP65
Electrical:
Li-ion rechargeable, built-in charger
autonomy: typical 10 hours
accepts primary batteries for
immediate availability
Dimensions: 320 X 145 X 175 mm
3
(12.59 x 5.70 x 6.88 cu in)
Weight: NaI version 4.9 kg (10.8 lb),
LaBr version 4.02 kg (8.86 lb)
ACCESSORIES
Carrying case, carrying strap
External main power supply
SPIR-Companion software
OPTIONS
Alpha – beta probe
Embedded GSM / GPRS / Edge / 3G
Desk and wall holder
www.mirion.com
145203EN-D
SPIR Companion software
Common to all libraries
Source search library
Nuclear accident library (*)
(*) Firmware version V1.2.3 and up
40K, 226Ra + daughters, 232Th + daughters
NORM
SNM
merged
SNM
expert
Other
Medical
Industrial
Nuclear
U, Pu, 237Np
U, LEU, HEU, 233U, HBPu, MBPu, LBPu,
238Pu, 241Pu, 237Np
Neutron, Bremsstrahlung (ie Beta), Unknown
18F,/511keV, 51Cr, 67Ga, 99mTc, 99Mo,
103Pd, 111In, 123I, 125I, 131I, 133Xe,
153Sm, 201Tl
22Na, 54Mn, 57Co, 60Co, 75Se, 133Ba,
137Cs, 152Eu, 154Eu, 166Ho, 207Bi, 228Th,
241Am
18F/511keV, 60Co, 106Rh, 131l, 132l, 133l,
Xe133, 134Cs, 137Cs, 152Eu, 154Eu