Manual

315 Daniel Zenker Drive
200 IST Center
Horseheads, NY 14845
USA
T +1.607.562.4369
+1.800.432.1478
F +1.607.562.4392
Station Road
Alton, Hampshire
GU34 2PZ, UK
T 01420.541600
F 01420.541700
Kaiser-Konrad-Str 93a
D 53225 Bonn
Deutschland
T 0228.625088
F 0228.626300
7.87”
(200mm)
7.87”
(200mm)
SPECIFICATIONS AND PERFORMANCE
PYROMETER SENSOR
Temperature Range 800°–2500° F (427°–1371°C)
Temperature Accuracy 1.0% Full scale
Measurement Technique Single wavelength pyrometry
Spectral Regions Narrow band centered at
1.6 microns
Spot Size 1/24 of the horizontal width
of the image
Spatial Scan Resolution 47 horizontal x 35 vertical width of the image
Scan Rate Scan speed varies with size & number of
TMZs or via operator adjustment
CAMERA
Power 115-230 VAC, 50/60 Hz
Camera Detector Solid state color image sensor
System Resolution 300 lines minimum throughout the image
Video 1.0V p-p, 75ohm, CCTV signal
/VTN: NTSC or /VTP: PAL video timing
selected at time of order
Control Iris adjustment on rear of camera; remote
iris adjustment from the processor
Application Filter Filters are provided to match your process
and maximize performance. Contact your
Sales Representative
ENCLOSURE
Construction /CEI: Corrosion-resistant, insulated, air-
cooled, NEMA 4; /CEW: Corrosion-resistant,
water-cooled, NEMA 4
Cooling Type Vortex cabinet cooler, 25 SCFM @ 100 psi
(13 dm3/sec @ 690 kPa); instrument-quality
air required or water cooled option available
Ambient Environment Max. 140°F (60°C) with negligible radiant heat
load. Water cooled option available to handle
high radiant heat environment
MECHANICAL
Video Output Jack Female PL-259 “UHF” type
Power Input Jack Removable waterproof miniplug (JOY type
TP, female 3-conductor; mating power
cord provided)
Enclosure Cooling Input 1/4” brass quick-disconnect nipple; mating
coupler (Snaptite BVHC4-4F) provided
Lens Cooling Input 1/2” brass quick-disconnect nipple; mating
coupler (Snaptite BVHC8-8F) provided
Weight 30 pounds (14kg) for standard air-cooled
conguration (lens and camera)
LENS
Construction Air or water-cooled 304 stainless steel
outer shroud. Sapphire window for max.
environmental protection
Type 45° offset lens
Length (L) /OAL30W553
27.7” (694mm)
For details on our full range of
Spyrometer lenses, please contact your
sales representative.
Field of View Wide - 75° H x 58° V;
Medium - 50° H x 38° V;
Narrow - 35° H x 26° V
Diameter /L: 1.5” (38mm); /OAL: 2.0” (51mm);
/WCL: 2.5” (63mm)
Cooling Requirements Instrument quality air, 25 40 SCFM
(12–19 dm3/sec) @ 5–15 psig
(34–103kPa) required for straight view lens
Thermocouple /TJ: Type J thermocouple option;
/TK: Type K thermocouple option
© Copyright 2008. All rights reserved for trademark and registered trademark
information.
The above specications may vary according to system conguration.
Mirion Technologies Incorporated reserves the right to amend or change the
information on this sheet without prior warning.