Owner's manual

Featuring:
radiation monitoring
systems
A Mirion Technologies Division
IM 201L
Particulate, Gas & Iodine Monitors
I M 201L
Iodine Monitor
The IM 201L monitor forms part of the RAMSYS
product line.
It has been developed to continuously measure
the gamma volumetric activity of radioactive iodine
sample, in both molecular and organic forms (methyl
iodine), contained in air drawn from stacks, ventilation
ducts or working areas. An NaI scintillation
detector faces the activated charcoal cartridge in
which Radioactive Iodine is trapped. The proximity of
the detector and the cartridge, enclose within a
4 π/5 cm (4 π/2 in) lead shielding, serves to optimize
detection ef ciency. A
241
Am radioactive source built
into the NaI christal allows compensation of tempera-
ture and aging related drifts. The spectrometry
capability, based on a 1024 channel spectrum
analysis, allows radio iodine isotope localization.
APPLICATIONS
Ef uent release monitoring
Stacks, etc.
Radioprotection of workers
Containment atmosphere
Control room air, etc.
RELATED MONITORS
IM 201M: mobile version
IM 201S: seismically quali ed version
PING 206S: version with particulate and noble gas
PIM 206S: version with particulate
FEATURES
Embedded
241
Am source for energy spectrum
stabilization against temperature changes and aging
1024 channels spectrum analysis
Ef uent trapping of both molecular and organic forms
of iodine

Summary of content (2 pages)