Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-dit Calès, route d’Eyguières
F-13113 Lamanon
France
T +33 4 90 59 59 59
F +33 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
www.mirion.com
151199EN-E
PHYSICAL CHARACTERISTICS
Compliant with IEC 61526Ed. 3, ANSI 42.20(*)
(*) isotropy
241
Am and
137
Cs with ± 75° angle
Measurement and display:
X and gamma energy range: 15 keV to 7 MeV
Energy response better than ± 20% (typically ± 10%)
from 16 keV to 7 MeV
Accuracy Hp(10):
≤ ± 10%* (
137
Cs, ~ 24 mSv/h); ≤ ± 15%*, (
241
Am, ~ 23 mSv/h);
≤ ± 19%** X-ray 16keV
(*including ± 5% extended uncertainty k=2) ;
(** including ± 9% extended uncertainty k=2)
Display units: mSv, μSv, or mrem
Display dose: 1 μSv to 10 Sv (0.1 mrem to 1000 rem)
Display rate: 10 μSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h) or
1 μSv/h to 10 Sv/h (0.1 mrem/h to 1000 rem/h)
extended option.
Measurement range: 1 μSv to 10 Sv (0.1 mrem to 1000 rem)
0.1 μSv/h to 20 Sv/h (0.01 mrem/h to 2000 rem/h),
Saturation indication above 10 Sv (1000 rem) or 10 Sv/h
(1000 rem/h)
Dose Rate Linearity:
< ± 20% up to 10 Sv/h (1000 rem/h) (Co and X H30 20 keV)
< ± 20% up to 6 Sv/h (600 rem/h) (Pulsed X-rays
20 ms width, 1, 10 & 20 pps)
ELECTRICAL CHARACTERISTICS
Standard AAA (LR03) 1.5 V Alkaline battery
9 calendar month battery life (typical, 8 h per day,
5 days per week in run mode, without excessive alarms)*
2500 h battery life in continuous run, without excessive alarm*
*0.2% of the time in alarm
MECHANICAL CHARACTERISTICS
Rugged, high impact polycarbonate-ABS case
Dimensions: 86 x 56 x 21 mm (3.4 x 2.2 x 0.8 in) without clip
Weight with alkaline battery and clip: < 88 g (3.1 oz)
Worn by a replaceable clips (2 different style back clips or one
a front-facing clip for DMC worn inside the pocket)
ENVIRONMENTAL CHARACTERISTICS
Temperature range: -10°C to 50°C (14°F to 122°F): deviation
in response less than ± 5%
Relative humidity: < 90% at 42°C (108°F)
Storage: -20°C to 71°C (-4°F to 160°F)
Shock, vibration and drop resistant (1.5 meter on concrete)
Waterproof IP67: 1m (39 in) during 1 hour
EMC: complies and exceeds standards by a large
margin (
(
(
(
(
(
(
compliant, certi cate number: 153720)
MIL STD 461F RS103 (pulsed electric  eld): exceeds 200 V/m
from 30 kHz to 5 GHz
MIL STD 461F RS101 (magnetic  eld 30 Hz - 100 kHz)
Factory calibration approved under ISO/IEC 17025,
COFRAC accreditation Nber 2-1663
Histogram Features
Dose increments with a 1 μSv (0.1 mrem) resolution saved on
non volatile memory (EEPROM) in con gurable steps
(10 s, 60 s, 10 min, 1 hour, 24 hours)
Dose rate saved at each dose increment when using
enhanced communication protocol
Event log (alarms, faults, changes) saved during the selected
time periods
Stores data for several consecutive workers entries and exits
(more than 1000 steps)
Display Features
Large LCD display with high quality white backlighting
8 alpha-numeric digit display for full name display
(scrolling) and  x dose/rate display format
3 top LEDs for Alarming (Red), gamma counting (Green)
and Hp(0.07) or Neutron counting (Blue)
2 push buttons for an easy customized data and
parameters display
Alarm Features and communication
Alarming speaker with level of 85 dB (A) typical
(> 90 dB (C) peak) at 30 cm (11.8 in), frequency < 4800 Hz
Vibrating alarm
High ef ciency red  ash LED on front
3 Top LEDs and display indicators
Dose/rate alarms, adjustable over the display range
Dose/rate warnings, adjustable over the
display range and acknowledgeable
Remaining time before alarm and run time alarms
Hands-free communication, frequency: 125 kHz
Backward compatibility with existing readers
(LDM 2000, LDM 220/210)
Range of communication with LDM 220: 10 cm (3.9 in)
Enhanced protocol to support additional features with the
new readers (LDM 320D, LDM 320W, LDM 2000 (*))
* requires reader  rmware upgrade
DMC 3000
Active Dosimetry - Dosimeters