BIM 201K
Irradiation Monitors
PHYSICAL CHARACTERISTICS
Radiation detected: beta
Detector: 2’’ thin plastic scintillator + PMT +
embedded LED
Energy range: > 30 keV
Typical measurement range: 3.7 10
+3
to 3.7 10
+9
Bq/m
3
(10
-7
to 10
-1
μCi/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature: +10°C to +40°C
(+50°F to +104°F)
Periodic temperature (processing unit):
-5°C to +55°C (+23°F to +131°F)
Periodic temperature (detector):
+0°C to +60°C (+32°F to +140°F)
MTBF (processing unit): > 50 000 hours
MTBF (detector): > 20 000 hours
TID: 100 Gy (10
+4
rad)
Protection index: IP65 and IK07
MECHANICAL CHARACTERISTICS
Dimensions:
Processing unit: 390 mm x 196 mm x 187 mm
(15.3 in x 7.7 in x 7.3 in)
Detector: 240 mm x 279 mm x 276 mm
(9.4 in x 11.9 in x 10.8 in)
Weight:
Processing unit: 4.5 kg or 7 kg (10 lb or 15.5 lb)
Detector: 11 kg (24.2 lb)
Color: gray RAL 7030 (decontaminable paint)
ELECTRICAL CHARACTERISTICS
Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING (Applicable to LPDU only)
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60761-1 and IEC60761-3
Environmental: IEE323 and IEC60780
Seismic: IEEE344 and IEC60980
EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4
VERSIONS
230 Vac or 120 Vac
LPDU or LPU
With or without RS485 junction box
With or without check source
With or without heater
Detector cable length: from 20 m (66 ft) to 65 m
(213 ft)
Junction box cable length: 5 m (16.4 ft) or 10 m (32.8 ft)
ACCESSORIES
Calibration tools
Software
Ethernet
USB converters
Seismic quali ed wall mounting brackets for LP(D)U
www.mirion.com