User's Manual Part 2

D-2 Specification
1
This specification has been verified in accordance with ASTM-E1112-00.
*Operation and storage outside the stated temperature and humidity range may degrade performance.
**Subjecting to shock outside the stated range may degrade performance.
Parameter Value Condition/Note
Functional Attributes
Temperature sensing range 25 °C to 50 °C Ingestible capsule
-20 °C to 60 °C Dermal patch
Temperature sensing accuracy ±0.10 °C 32 °C to 42 °C
(Guaranteed
1
)
±0.05 °C 32 °C to 42 °C
(Typical)
±0.25 °C -20 °C to 32 °C
(Guaranteed)
±0.25 °C 42 °C to 60 °C
(Guaranteed)
Temperature display resolution ±0.01 °C
Display update rate 15 seconds Average
Monitor battery life 10 days (240 hours) with
10 sensors on line, plus 20
days standby
Battery life increases with
fewer sensors on line
Sensor battery life 1 year storage plus 10
days active transmission
Capsule and patch
Calibration None required
Number of co-active sensors One to ten Per monitor
Sensor identification Automatic tracking
Crosstalk Not allowed
Maximum logging time 6 days
10 days
10 sensors
5 sensors
Environmental Attributes
Moisture protection IEC529-IP52
NEMA 250-5.3
Monitor and sensors
Storage temperature* -20 to 50 °C @ 5-95% humidity
Monitor and sensors
Operating temperature* 0 to 40 °C Monitor
Shock** 1 meter drop to tiled
concrete floor
Monitor
Transportation Environment
Attributes
Moisture protection IEC529-IP52
NEMA 250-5.3
Monitor and sensors
Storage temperature -20 to 50 °C @ 5-95% humidity
Monitor
Shock 1 meter drop to tiled
concrete floor
Monitor