User's Manual Part 2
5-13
Irregular Rhythm Responses
Per AAMI EC13-5.1.2.1(e)
Regulatory Standards
VitalSense-XHR Sensor will be tested according to the following
standards:
ECG Complex Waveform VitalSense-XHR
Response
(Avg BPM)
Expected
(BMP)
3a: Ventricular bigeminy AAMI 3a 83 80
3b: Slow alternating ventricular
bigeminy
AAMI 3b 58 60
3c: Rapid alternating ventricular
bigeminy
AAMI 3c 58 120
3d: Bidirectional systoles AAMI 3d 123 90
Test Standard Description VS-
XHR
Sensor
Multi-
charger
Reader/
Charger
IEC60601-1 Medical Electrical Equipment - Part 1:
General Requirements for Safety
3
IEC60601-1-2 Medical Electrical Equipment - Part 1-2:
General Requirements for Safety - Collateral
Standard: Electromagnetic Compatibility -
Requirements and Tests
3 3 3
IEC60950-1 Information Technology Equipment - Safety -
Part 1: General Requirements
3 3
47 CFR 15.229 Intentional Radiator Permissive Change
3
ASTM/AAMI-
ESI:1993
Safe current limits for electromedical
apparatus
3
AAMI-EC13:2002 Cardiac Monitors, Heart Rate Meters, and
Alarms
3