User's Manual Part 2

5-10 VS-XHR Sensor
VitalSense XHR Sensor Specification
Intended Use
VS-XHR Sensor is intended to be used as a heart rate and/or respiration
rate monitor, in conjunction with the VitalSense Integrated Physiological
Monitoring System. VS-XHR Sensor is not an ECG monitor. VS-XHR
Sensor is not equipped with alarms to signal tachycardia or any other
cardiac arrhythmias.
Physical Attributes:Sensor
ECG Electrode Types
VS-XHR Sensor is considered a two-lead ECG detection sensor. It uses a
modified Lead I or Lead II connection. See Site Preparation ” on
page5-7 for a description of the preferred lead placement sites.
Parameter Value Condition/Note
Size 195 mm length, overall 38 mm diameter,
12.5 mm tail clip
Outer dimensions
Weight 12 grams With no ECG
electrodes attached
Case material Polycarbonate/ABS Flammability rating
94V-0
Attachment clip
material
Stainless steel
Attachment type Adult ECG snap
Battery type 3.0 volt lithium rechargeable Not user replaceable
Indicators Green LED Coincident with heart
beat
Lead Type Sensor Body Sensor Tail
Lead I White (RA) Red (LA)
Lead II White (RA) Red (LL)