User's Manual Part 1

2-10 Monitor Operation and Sensor Activation
Administration
Both the Capsule Sensor and the Dermal Patch contain ferromag-
netic materials incompatible with MRI (magnetic resonance imag-
ing) and should not be worn nor ingested if an MRI is planned. Any
subject ingesting or affixing the above sensors is to wear an MRI
Warning wrist band (supplied with each sensor) as long as the sen-
sors are in place.
Dermal Patch Sensor
1 Remove excess hair from the area in which the sensor will be mounted.
2 Prepare the skin by cleaning with an alcohol-moistened wipe.
3 Following successful activation, use the tab to peel away the protective
layer.
4 Affix the Dermal Patch Sensor to the skin.
5 Affix an MRI Warning wrist band (supplied) to the subject with
instructions that it is to be worn as long as the Dermal Patch is in place.
Keep the VitalSense Monitor within 2 meters of your body.
NOTE: Avoid the use of lotions and emollients in the area of the
Dermal Patch Sensor. Their uses may cause the adhesive to loosen.
Capsule Sensor
Do not chew. There are metal components inside the capsule. Swal-
low whole with liquid.
1 Following successful activation, swallow the Capsule Sensor with
approximately 8-ounces of water, juice, or soft drink.
2 Keep the VitalSense Monitor within 1 meter of your body.
3 Affix an MRI Warning wrist band (supplied) to the subject with
instructions that it is to be worn until the Capsule Sensor has been
expelled.