User's Manual
D-4 Specifications
Specication for Actiwatch-Score
ENVIRONMENTAL
Operating Storage
Temperature
0° C to 40° C
(32° F to 104° F)
-10° C to 50° C
(14° F to 122° F)
Relative Humidity
0 to 95% RH 0 to 95% RH
Atmospheric Pressure 500 to 1060kPa
(5600 feet to sea level)
PHYSICAL
Dimensions:
37 L x 35 W x 12 H (mm)
Weight:
19 grams w/o band
25 grams with standard band
Case Material: Glass-filled Nylon
Frame and Battery Cover: Titanium
Standard Wrist Band: Nylon with buckle
ELECTRICAL
Battery Type: CR 2025
Battery Life: 3 Months Typical
Protection Against Electric
Shock:
Internally powered
Degree of Protection Against
Electric Shock:
Type BF Applied Part
Modes of Operation: Continuous
Electromagnetic
Compatibility:
The device meets the
requirements of EN 60601-1-2,
second edition (2001).
Fuses: There are no user-replaceable
fuses.
Degree of Protection Against
Ingress of Water:
Dripping and light splashing of
liquids
IEC Standard 60529 IPX2