User's Manual

Monitoring ECG Understanding the ECG Display
TM80/TM70 Telemetry Monitor Operator’s Manual 7 - 17
[Auto]: the TM80/TM70 will use the heart rate from the ECG measurements
as the alarm source whenever a valid heart rate is available. If the heart rate
becomes unavailable, for example the ECG module becomes disconnected,
the TM80/TM70 will automatically switch to PR as the alarm source.
[Both]: both HR and PR are used as the alarm source for HR/PR.
7.6 Understanding the ECG Display
7.6.1 HR Digital Area
The HR digital area displays:
1. Parameter name
2. Measurement unit
3. Heart rate value
4. PR source: when ECG leads are not connected and valid PR value is detected, PR
source is displayed.
5. Activation state off symbol: when HR/PR alarm is switched on and valid HR/PR
value is detected, HR/PR alarm high and low limits in place of the activation state
off symbol are displayed.
7.6.2 About the HR Digital Area
The HR area displays heart rate in the unit of bpm with a resolution of 1 bpm.
If the HR measurement is invalid, “- - -” displays in place of the HR value.
The HR value displays “0”, when the HR value is less than 15 bpm.
When lethal arrhythmia alarms are switched off, corresponding alarm off message
is displayed in the HR digital area.
NOTE
When [Alm Source] is set to [HR], systole beep comes from heart beat.
When [Alm Source] is set to [PR], systole beep comes from pulse rate.
1 2
3
5
4