7 - 2 TM80/TM70 Telemetry Monitor Operator’s Manual
7.1 Introduction
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it
on the monitor as a waveform and a numeric.
ECG monitoring provides 3-, 5-, and 6-lead ECG monitoring, ST-segment analysis,
arrhythmia analysis, and QT/QTc measurements.
Operations such as configuring QRS threshold, adjusting ST point/ISO point/J point, set-
ting ST template/QT template are performed at the CentralStation. For details about
these operations, refer to Chapter 12 Monitoring with the TM80/TM70 at the CMS.
7.2 Safety
WARNING
This equipment is not suitable for direct cardiac application.
Use manufacturer specified electrodes and lead wires.
Make sure the conductive parts of electrodes and associated connectors
for applied parts, including the neutral electrode, do not contact any
other conductive parts including earth.
Periodically inspect the electrode application site to ensure skin quality.
If the skin quality changes, replace the electrodes or change the applica-
tion site.
Use defibrillation-proof ECG leadwires during defibrillation.
Do not touch the patient or any device connected to the patient, includ-
ing the bed and gurney, during defibrillation. Otherwise serious injury
or death could result.
CAUTION
Interference from a non-grounded instrument near the patient and elec-
tro-surgery interference can cause problems with the waveform.