User's Manual
Pausing Alarms Alarms
6 - 8 TM80/TM70 Telemetry Monitor Operator’s Manual
6.7.3 Restoring the Default Alarm Settings
1. Press the key to enter the main menu.
2. Tap [Alarms].
3. Tap [Alarm Limits]. The [Alarm Limits] menu is displayed.
4. Tap [Defaults].
5. In the dialog that pops up, tap [OK].
6.8 Pausing Alarms
When an alarm occurs, follow this procedure to pause the alarm system.
■ Press the key to enter the main menu, and then tap [Alarm Pause] from the
[Commands] section.
OR
1. Flick your finger up at the bottom of the main screen to display the quick keys
area.
2. Tap the [Alarm Pause] quick key.
When alarms are paused, the following indicators are given:
■ No physiological alarm will be presented within the alarm pause time.
■ For technical alarms, alarm sounds are paused, but the alarm lamp and alarm
messages remain presented.
■ The remaining alarm pause time is displayed in the message area.
■ The alarm pause symbol is displayed in the upper right corner of the main
screen.
NIBP NIBP-S
(mmHg)
(SYS×0.68 +
10)
SYS×0.86 + 38
Adult: 45~270
Pediatric: 45~185
NIBP-D
(mmHg)
(Dia×0.68 + 6) (Dia×0.86 +
32)
Adult: 25~225
Pediatric: 25~150
NIBP-M
(mmHg)
(Mean×0.68 +
8)
Mean×0.86 +
35
Adult: 30~245
Pediatric: 30~165
Module Parameter
Lower Limit Upper Limit
Auto Limit Range
Adult/
Pediatric
Adult/
Pediatric