User's Manual

Pausing Alarms Alarms
6 - 8 TM80/TM70 Telemetry Monitor Operator’s Manual
6.7.3 Restoring the Default Alarm Settings
1. Press the key to enter the main menu.
2. Tap [Alarms].
3. Tap [Alarm Limits]. The [Alarm Limits] menu is displayed.
4. Tap [Defaults].
5. In the dialog that pops up, tap [OK].
6.8 Pausing Alarms
When an alarm occurs, follow this procedure to pause the alarm system.
Press the key to enter the main menu, and then tap [Alarm Pause] from the
[Commands] section.
OR
1. Flick your finger up at the bottom of the main screen to display the quick keys
area.
2. Tap the [Alarm Pause] quick key.
When alarms are paused, the following indicators are given:
No physiological alarm will be presented within the alarm pause time.
For technical alarms, alarm sounds are paused, but the alarm lamp and alarm
messages remain presented.
The remaining alarm pause time is displayed in the message area.
The alarm pause symbol is displayed in the upper right corner of the main
screen.
NIBP NIBP-S
(mmHg)
(SYS×0.68
10)
SYS×0.86 38
Adult: 45~270
Pediatric: 45~185
NIBP-D
(mmHg)
(Dia×0.68 6) (Dia×0.86
32)
Adult: 25~225
Pediatric: 25~150
NIBP-M
(mmHg)
(Mean×0.68
8)
Mean×0.86
35
Adult: 30~245
Pediatric: 30~165
Module Parameter
Lower Limit Upper Limit
Auto Limit Range
Adult/
Pediatric
Adult/