Configuring the TM80/TM70 Maintenance Menu
TM80/TM70 Telemetry Monitor Operator’s Manual 13 - 5
3. Tap to exit the [Alarms] menu.
Section & Options Description Settings*
Sounds
Style Allows an authorized user to set the alarm tone
pattern.
ISO, Mode 1, Mode 2
Minimum Alarm Vol-
ume
Allows an authorized user to configure the
minimum alarm volume.
The minimum alarm volume defines the mini-
mum value you can set for the alarm volume.
For example:
If the minimum alarm volume is set to [5], the
minimum value you can set for the alarm vol-
ume in the [Audio Volume] menu is 5 (as
shown in the following figure).
If the minimum alarm volume is set to [Off] and
the [Sounds] is set to 0, the alarm sound is
turned off and the symbol appears on the
screen.
Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
Timeout
Reminder Tone Allows an authorized user to enable or disable
the reminder tone.
If the alarm sound is turned off, enabling this
setting can issue a periodic reminder tone.
On, Off
Reminder Interval Allows an authorized user to configure the
intervals between the alarm tones.
1 min, 2 min, 3 min, 5 min, 10 min
Technical Alarm Priority
ECG Lead Off Allows an authorized user to configure the
alarm priority.
Low, Medium, High
SpO
2
Sensor Off Allows an authorized user to configure the
alarm priority.
Low, Medium, High
* The factory default settings are in bold.