Operator's Manual

Maintenance Menu Configuring the TM80/TM70
13 - 4 TM80/TM70 Telemetry Monitor Operator’s Manual
13.2.2.1 Configuring Device Location
You can change the hospital name and set whether to allow modifying a department
name, room number, and bed number in the [Patient Info] menu. To do so, follow this
procedure:
1. In the [Maintenance] menu, tap [General].
2. Tap [Device Location].
3. Tap [Facility].
4. Enter the desired name.
5. Tap [Department ].
6. Select the desired option.
[Fixed]: you cannot modify [Department] in the [Patient Info] menu.
[Department] is the one entered in the [Department] field of the [Device
Location] menu.
[Unfixed]: you can modify [Department in the [Patient Info] menu.
7. Tap [Bed No./Room No].
8. Select the desired option.
[Fixed]: you cannot modify [Room No.] and [Bed No.] in the [Patient Info]
menu. [Room No.] and [Bed No.] are the ones entered in the [Room No.]
and [Bed No.] fields of the [Device Location] menu.
[Unfixed]: you can modify [[Room No.] and [Bed No.] in the [Patient Info]
menu.
13.2.3 Configuring the Alarms Menu
Select [Alarms] to configure the alarm tone, reminder tone, reminder interval, minimum
alarm volume, and technical alarm priority.
1. In the [Maintenance] menu, tap [Alarms].
The current setting displays to the right of the option.
2. Select the options described in the following table.
NOTE
If [Department] and [Bed No./Room No.] are set to [Unfixed],
Department name, Room No., and Bed No. are cleared each time you
discharge a patient.